Robotic Scanning Device for Intraoperative Thyroid Gland Endomicroscopy.
| Author | |
|---|---|
| Abstract | 
   :  
              Probe-based confocal laser endomicroscopy has a high potential to be a promising tool that can provide intraoperative high-resolution in vivo morphological imaging at cellular and subcellular levels for thyroidectomy, and allow real-time assessment of tumor margins. However, the typical images acquired with this technique cover a very small area limited by the field of view of the probe, accompanied by tissue deformation and inconsistent probe-tissue contact when operated manually. In this paper, a novel compact robotic device for large area scanning has been developed. The device can scan a large surface in a spiral trajectory by rotating the tip frame along the spiral groove of the base frame. The fiber Bragg grating sensor with a passive linear structure is used to detect and maintain a stable probe-tissue contact force during scanning. An active linear actuation is also integrated for adjusting the probe-tissue contact level prior to each scan. Results demonstrate that the scanning device ensures a suitable probe-tissue contact force and compensates for simulated hand tremor. Mosaicing results of lens tissue paper and porcine belly tissue with both bench and hand-held experiments show the effectiveness and usability of the device, demonstrating the potential clinical value of the system.  | 
        
| Year of Publication | 
   :  
              2018 
           | 
        
| Journal | 
   :  
              Annals of biomedical engineering 
           | 
        
| Date Published | 
   :  
              2018 
           | 
        
| ISSN Number | 
   :  
              0090-6964 
           | 
        
| URL | 
   :  
              https://dx.doi.org/10.1007/s10439-018-1978-x 
           | 
        
| DOI | 
   :  
              10.1007/s10439-018-1978-x 
           | 
        
| Short Title | 
   :  
              Ann Biomed Eng 
           | 
        
| Download citation |